A ground loop forms when a measurement circuit is connected to earth at two or more points. The earth's potential varies from point to point, so current flows between those connections. When that current passes through a signal wire, or through the reference of a pH sensor, the reading becomes noisy and offset. pH and ORP sensors are the most affected, because they measure very small voltages: just 59 mV per pH unit.
Download the electrical noise & interference guide (PDF)
What a ground loop does to a pH reading
The classic symptom is a sensor that reads correctly in buffers but gives a reading grossly in error in the process liquid. In a typical installation the pH sensor is connected to earth through the process liquid and piping. If the circuitry in the pH analyser becomes connected to a second earth, current flows through the reference electrode. That develops a voltage in series with the other cell voltages, so the pH reading moves well away from the expected value. Ground loop currents are often unstable, so the reading is often noisy too.

How to check for a ground loop
Suspect a ground loop if the system starts acting strangely or erratically, or if the reading changes when you touch a cable or move the sensor. Changes such as adding a motor or agitator, or working on any electrical item, can upset the balance, so check again after them.
- Remove the pH sensor from the process liquid.
- Calibrate the sensor in buffers. Make sure there's no direct electrical connection between the buffer container and the process liquid or piping.
- Strip back the ends of a heavy gauge wire.
- Connect one end of the wire to the process piping, or better, place it in the process liquid. Place the other end in the container with the buffer and sensor.
- If the reading changes or becomes noisy once the wire is connected, a ground loop exists.
- If no symptoms develop, a ground loop probably doesn't exist.
Ground loops damage the sensor
A ground loop doesn't only distort the reading: it polarises the sensor and damages the reference. The polarisation can cause wrong readings even after the sensor is removed from the ground loop. It may fade with time, but recalibration may be needed. Left in place, a ground loop will destroy the sensor's responsiveness and cause premature failure, so act on it as soon as you find it.
Could it be a faulty sensor?
No. pH and ORP sensors are passive electrodes: they can't generate or store current, so any extra millivolts must be coming from their surroundings. If a sensor calibrates and reads correctly in buffer, it is working correctly. Nothing can be done to the sensor to correct the reading, other than removing the source of interference. If you're concerned about a sensor's health or response, see the sensor health and performance guide.
Other sources of electrical noise
- AC power lines: especially where sensor cables run parallel to high-voltage or high-current lines.
- Electromagnetic interference (EMI): from motors, transformers and radio transmitters, coupled into the sensor wiring and electronics.
- Radio frequency interference (RFI): from radio and microwave sources.
- Nearby electrical equipment: variable frequency drives, inverters and large motors.
- Capacitive coupling: between conductive objects close together at different voltages.
- Improper shielding or wiring: poorly shielded or grounded sensor cables pick up ambient noise.
- Chemical interference: not electrical, but the presence or absence of other ions affects the sensor. Low-conductivity solutions in particular can give very unstable, noisy pH and ORP readings.
Careful installation prevents most of these: proper shielding and grounding, twisted pair cables for signal wires, and sensor wiring kept away from high-voltage or high-current power lines. Check routinely for corrosion and loose connections.
How to track down the source
A ground loop is a problem in the electrical system, not a fault in the equipment. Turtle Tough can't advise on modifying your electrical system: work with qualified electrical personnel who know your installation. The steps below are a general starting point, not a definitive list.
Earth the installation
All electrical equipment must be properly earthed, and sensors installed in adequately earthed environments. Where possible, install the sensor in a metallic fitting such as a metal immersion rod that is properly earthed to a common earth, which gives stray current a path. In poly pipe or plastic fittings, find another way to earth the installation to divert stray current before it reaches the sensor. For inline plastic pipework, an earthing sensor can be a good short-term solution, while you locate and isolate the source.
Isolate the sensor from the power and signal side
- For digital sensors, use the Handheld Field Communicator (HFC) or the Windows interface. The HFC powers the sensor from its own 9 V battery, isolating it completely from your electrical installation.
- Without an HFC, disconnect the analyser's outputs, then power the analyser from a stand-alone battery.
- Or use a Turtle Tough isolation module, which provides galvanic and optical isolation. We recommend it for every self-integration installation.
- Ground loops can also come through the 4–20 mA output to the PLC if the PLC has poor isolation. Temporarily disconnect the PLC to rule it out.
Find the equipment on the process side
- Keep the pH sensor at least 50 cm from active sensors such as DO, conductivity or level sensors, which generate current near the tip. Allow more distance in vessels that aren't well earthed, such as plastic or rubber-lined vessels.
- Keep the sensor as far as possible from high-current equipment such as pumps, motors and variable speed drives. These are the primary suspects: if it isn't a poor earth, it's often poor isolation of their power supplies or insufficient shielding.
- Switch off one device at a time and note when the problem goes away. This is the quickest way to find the source on the process side.
- For intermittent problems, enable data logging (locally or through your PLC). Look for periods when the reading returns to normal, and check whether they line up with equipment being switched off. Or look for when abnormal readings coincide with something switching on, such as a pump or motor.
If you can't remove the cause: isolate
Signal and power isolators break the galvanic path (DC continuity) between grounds while letting the analogue signal through, and can also remove AC common-mode noise. Whatever method you choose, the isolator must provide input, output and power isolation. Without all three, a new ground loop can form between the isolator's power supply and the input or output signal. As a general rule: isolate, isolate, isolate, and always make sure the analyser is properly earthed to the installation.
Preventing ground loops in new installations
It's much easier to avoid ground loops at the planning and installation stage than to diagnose them in the field. Things to consider:
- Grounds are often not the same and are often separated by distance, so their potentials are not equal.
- The loop isn't always in the 4–20 mA signal: check non-isolated RS-485 signal wires and non-isolated power, output and input grounds.
- Ground potential differences come from noise, the resistance of the ground path, and poor initial power rail installation.
In a larger network, use a dedicated instrumentation system ground bus, and connect the signal common, cabinet ground and instrumentation AC power ground to it. The bus is tied to earth through the building ground and plant ground grid. Vendor-supplied instrumentation cabinets usually contain a DC signal common bus and a power supply common bus, tied together at a master ground bus. The cabinet ground is a safety ground that protects people and equipment from shock and drains static and EMI; it stays separate from the DC signal ground until the master ground bus. The AC service ground is a single-point termination of the system AC power, at the neutral-to-ground bond of the main AC power isolation transformer, and terminates at a single point on the plant ground grid (usually the grounding electrode).
More on installing sensors: installation recommendations.
Frequently asked questions
Why does my pH sensor read correctly in buffer but wrong in the process?
That's the classic sign of a ground loop: the circuit is earthed at two points with different potentials, so current flows through the sensor's reference and shifts the reading, often making it noisy too.
To check: calibrate the sensor in buffer in a container with no electrical connection to the process, then run a heavy-gauge wire from the process liquid (or piping) into the buffer. If the reading changes or turns noisy, you have a ground loop.
Act quickly, as ground loops polarise and eventually destroy the sensor. Remove unintended ground connections, and if you can't, fit a signal isolator with input, output and power isolation. More in what is a ground loop?
Why are my pH or ORP readings jumpy or erratic?
Usually stray current or static, not a faulty sensor. pH and ORP sensors measure tiny voltages (just 59 mV per pH unit), so a small current can shift the reading a lot. Common causes are poorly earthed or shielded equipment, plastic pipework and plastic immersion rods, and nearby pumps, motors or variable speed drives.
Install sensors in metal fittings that can be properly earthed (stainless steel immersion rods rather than plastic), earth plastic pipework, and keep sensors as far as possible from high-current equipment. See installation recommendations.
This information is general in nature and isn't professional advice for your installation. Questions? Call +61 3 9872 5055, email , or tell us about your application.