Unsure about your sensor's health and performance?
This guide will show you how to calculate and use the sensor slope & offset to evaluate the sensor's health.
Using sensor slope to evaluate sensor health
A theoretically perfect sensor would have a slope of -59mV per pH unit. When delivered new, a sensor will have an actual slope of somewhere between 56-59mV. As the sensor deteriorates, so does its range per pH unit; accordingly the slope falls.
| Slope | Status | Description | |
| 56-59 mV | As new Fast response | Sensors in this slope range will be very responsive and accurate. They will respond to buffer solutions in a couple of seconds and typically stabilise in less than 30 seconds for calibration purposes. | |
| 50-55 mV | Good Moderate response | Sensors in this slope range will be moderately responsive. The sensor has started to age and more frequent cleaning and calibration may be required to maintain accuracy. Response time to buffers is now likely to be around 30 seconds and up to 2 minutes to stabilise for calibration purposes. | |
| 45-50 mV | Close to expiry Slow response | Sensors in this slope range will be slow to respond but for sites wanting to get the maximum lifespan from their sensors, they will still have some useful service life left. The sensor is now quite deteriorated and may require more frequent cleaning and calibration to maintain the required accuracy. Response time to buffers is now likely to be around 1-2 minutes, with up to 5 minutes to stabilise for calibration purposes. | |
| <45 mV | Expired Replace sensor | We would recommend not using sensors with slopes of less than 45mV. They will be extremely slow to respond and the accuracy will be low. |
Using sensor offset to evaluate sensor health
Sensor offset is another great indicator of sensor health. With poisoning and deterioration of the reference, the offset will begin to shift away from the starting point. When using the offset value as a performance indicator it is important to know what kind of sensor we are dealing with. When evaluating lab electrodes or sensors with very low impedances it is common to compare the offset to the theoretical zero point 0 mV, which is the signal that should be generated when placed in buffer pH 7. However, for industrial sensors using 0 mV as a baseline for sensor health is not recommended. Due to the nature of construction, many Turtle Tough sensors will have offsets from new that are significantly higher than your average sensor and further away from zero. High impedance components such as special glasses or reference junctions can have a significant effect on the sensors offsets. The offset limit for Turtle Tough digital smart sensors is +/- 250mV, so they can easily handle higher offsets than your average sensor and a high offset from 0 mV does not indicate that the sensor is in poor health. As such, when evaluating a Turtle Tough sensor it is important to note the CHANGE in offset from the new condition and record this as your baseline, rather than comparing it to 0 mV. The following table indicates sensor health as a change in offset from the starting condition of the sensor (ie how it was received from the factory).
| Change in Offset | Status | Description | |
| +/- 10 mV | As new Excellent Condition | Sensors that have shifted less than 10mv from new, indicate they are near or as new.
| |
| +/- 10 to 20 mV | Good Good Condition | Sensors with this change in offset range are now showing signs of deterioration but should continue to perform well. | |
| +/- 20 to 30 mV | Significant Deterioration | Sensors with this change in offset range are now showing significant deterioration. Provided that you are using an appropriate analyser, this offset is compensated for in the calibration procedure and you should continue to get service life from the sensor. | |
| +/- 40 mV or greater | Close to expiry | Sensors with this change in offset range are now close to expiry and replacement should be considered. On some analysers, calibration may not be possible. | |
| +/- 100 to 999 mV | Danger/Warning | Sensors with this change in offset range are experiencing severe damage to the reference system. This can often be caused by a stray electrical current within the installation or a severe chemical attack. This condition should be identified and rectified immediately. |
Notes
- This is a general guide only. Each site must determine an appropriate performance regime and sensor replacement criteria. If you have a critical process where very high accuracy is required you may choose to discard a sensor at a 50mV slope.
- Sensor health is a dynamic combination of both slope and offset. The analyser will compensate for both slope and offset during a calibration procedure which will determine the sensor’s overall performance.
- Response times and stability of the sensor will vary depending on your sensor and build type. Turtle Tough sensors can contain different sensing components depending on your application and this will affect response times.
- It is important to clean your sensor each time you calibrate, even if it does not look dirty. Poisoning and fouling of the reference system occur at the molecular level (invisible to the naked eye) and soaking in the appropriate cleaning solution will dissolve or leach out these contaminants. This will greatly extend sensor life and performance. Too often people ignore the cleaning requirement during the early stages until the sensor shows decreased performance. Often by this time, the damage is already done and a sensor will degrade much more quickly if good practices aren’t adhered to early on. Please refer to Turtle Tough’s sensor care.
- If you have significant deterioration of the offset, ie greater than 100mV this may indicate a serious problem with your installation and it is recommended you contact your Turtle Tough representative for further advice.
How do I calculate my slope?
Many instruments will automatically calculate the slope for you and report this value at the end of the calibration procedure. Often this information is stored in a diagnostic menu on your transmitter. You can also calculate it manually anytime you do a calibration, just by recording the mV values output by the sensor. Most transmitters will actively display the mV values as you calibrate. Consult your manufacturer manual to locate this value.
Steps
- Place the sensor in buffer 7 (zero point). Theoretically, you should get a value of 0. Don’t be alarmed if you don’t get exactly zero, as you very rarely will. Remember this is a theoretical number. Record your mV reading.
- Place the sensor in buffer 4. Theoretically, you should get a value of 177mV. Again, your actual number might vary. Record your mV reading.
- Now do the math. From buffer 7 to 4 we have 3 pH units and a total span of 177mV. To find our slope we divide our span 177mV by the 3 pH units = 59 mV (a perfect slope).
Notes
- Now, remember your real-world values will be different from theoretical. For example, in buffer 7 you may get 10mV reading and in buffer 4 you may get 174. This span would be 164mV across 3 pH unit which equals a slope of 54.67.
- Some people reference a percentage of slope. For example, in the above scenario, our sensor with a span of 164mV compared to the ideal 177mV is expressed as 164/177= 0.9265 or 92.65%.
How do I calculate my offset?
Simply place your sensor in buffer 7 and read the mV value displayed on the analyser.
Note
- You may need to access a submenu/function button to view the raw mV values. Buffer 7 (neutral) should generate the theoretical value of 0mV. The actual value that your sensor is reading in buffer 7 is the offset value.