Nernst equation — pH electrode mV response

A glass pH electrode generates a voltage proportional to the hydrogen ion activity in solution. At 25 °C, each pH unit produces a 59.16 mV change, crossing zero at pH 7 (neutral).

Isopotential point
pH 7 = 0 mV
Zero-crossing / neutral
Nernst slope (25 °C)
59.16 mV/pH
Theoretical ideal at 25 °C
Full-scale range
+414 to −414 mV
pH 0 to pH 14
25 °C | 59.16 mV/pH
+414 +296 +177 +59 0 −59 −177 −296 −414 Electrode output (mV) 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 pH value ACIDIC NEUTRAL ALKALINE pH 4.01 buffer +177 mV Isopotential point pH 7.00 = 0 mV pH 10.01 buffer −177 mV 59.16 mV per pH unit
Hover or tap on chart to read values
E = E° − (RT / nF) × ln(aH+)
E = −59.16 × (pH − 7) mV  at 25 °C

Analysers

TurtleTough TT MA RH 

Extreme sensors call for extreme analysers and the Turtle Tough range has been designed with exactly that in mind. Whatever the application, we have a range of Smart analysers and control options to interface with your DSS sensors. Click to view our range...

Interface Hardware

TurtleTough TT MA RH 

We have a range of Digital Smart System accesories to interface, calibrate and communicate with your DSS sensors. Click to view our range...